HMG 3000 Page 25
Edition 11/2007 - V02 R01-S02 Stock no.: 669712
If you have selected counter pulses as the
input signal, and have returned to the normal
measurement value display, you will find on
the far right of the function bar a symbol, as
shown in the screenshot. The counter is
currently stopped. If you select the symbol
using the arrow key and confirm with OK, the
counter is reset and will start again. To stop
it, select the symbol again using the arrow
keys and confirm again with OK.
In addition to the input channels (channels A
to channel J) there are also the virtual
channels Channel K and Channel L
available. These two channels can be used
to display a differential between 2 input
channels, or to calculate a power.
When setting to Differential the HMG 3000
calculates for channel K the difference
between channel A/channel B and for
channel L the difference between
channel C/channel D. To measure the
differential, sensors with the same
measurement range, unit (of measurement)
and number of decimal places must be used
on the input channels. These parameters are
then automatically adapted to channel K or L,
i.e. no further settings are necessary.
If you would like to measure Power (P*Q),
via channel K, the pressure must be
measured at channel A and the flow rate
must be measured at channel B. To
measure the power via channel L you have to
measure the pressure at channel C and the
flow rate at channel D. You can then select
kW, PS or hp as the unit (of measurement).
Entering the upper measurement range
merely limits the scale of the Y-axis in the
graphical display.