Automatic mode
In this menu point, you define whether the
recording (using the settings previously set)
is to be started just once or at regular inter-
vals.
Deactivated means that the recording is
started as predefined, and is ended once the
stop condition occurs and is saved.
Activated means that once the stop condi-
tion occurs, the current recording is ended
and saved, and a new recording is started
immediately with the same conditions. When
the stop condition occurs again, it is saved
under a consecutive number.
This continues until the (automatic) recording
is ended with Autostop and subsequent
stop condition or the memory is full.
The automatic operation reverts automati-
cally to "deactivated" each time a recording
is stopped, and must be "activated" again as
required.
Zoomed area
With the "zoomed area", an additional dis-
play area can be created during the re-
cording in the graphic mode.
I.e. when setting Zoom the x-axis is scaled
to the time span preset in this menu point (5
seconds / 10 seconds).
(The setting options for the zoomed area are
dependent on the preset sampling rate and
the specified recording time.)