HMG 3010 Page 46
Edition 11/2011 - V04 R01 Part No.: 669855
Event: Leaving measurement window
This trigger event is very similar to Specific
value. You can specify a Lower Limit and
an Upper Limit. The trigger is activated
when the measured value enters the meas-
urement window. However, recording is not
started until the measured value leaves the
measurement window in a rising or falling
direction.
Note:
Sub-channel setting only required for SMART sen-
sors (see Section 6.4) and CAN function (see Sec-
tion 6.5).
Event: Timed
The event setting using a time function
(Timed) is useful for measurements which
are to be repeated at specific intervals. In the
Settings menu you can specify the number
of Repetitions (max. 30,000), the Interval
and the Starting time. In the Start item you
can specify whether this event is to be acti-
vated immediately (Start immediately) or
not until after a certain Start date and Start
time.
In the
– Event: Timed –
example shown on the right, a measurement
is started 10 times every 30 minutes. The
trigger becomes active for the first time on 21
October 2006 at 14:20.
Up to 4 events can be specified for recording
triggered measurements. After making each
of the settings, return to the Trigger Set-
tings menu via Back. The menu now pro-
vides an overview of all the events and set-
tings for you to check.