HMG 3010 Page 48
Edition 11/2011 - V04 R01 Part No.: 669855
The next setting parameter for recording a
triggered measurement curve is Pre/post
trigger. After starting a triggered measure-
ment curve the HMG begins to record the
measured values for the recording time set
(e.g. 30 seconds). If the trigger event does
not occur during this time, the oldest meas-
ured values are replaced with new measured
values in the measured value memory. This
means that you always have the set re-
cording time as the "past" in the measured
value memory.
The Pre/post trigger option enables you to
specify how much of this "past" is to remain
in memory when the event occurs and how
much "future" you would like to record after
the event.
A Pre/post trigger setting of 50 / 50%
means that for a recording time of 30 s the
"oldest 15 s" are discarded, 15 s remain be-
fore the occurrence of the event, and re-
cording is done for another 15 s after the
trigger event occurs.
With Automatic mode you define whether
you wish to make just one recording (Deac-
tivated) or whether, after recording one
measurement curve, another curve is to be
recorded using the same settings (Acti-
vated). For a Simple triggered measure-
ment, select the Deactivated automatic
measurement option.
Zoomed area
With the "zoomed area", an additional dis-
play area can be created during the re-
cording in the graphic mode.
In other words, with the Magnifier setting,
the x-axis is scaled to the time span preset in
this menu point (5 seconds / 10 seconds).
(The setting options for the zoomed area are
dependent on the preset sampling rate and
the specified recording time.)